Access Immunoassay Systems Assay Protocol: Dil-AFP.

Recall: Z-1388-05 · Initiated July 18, 2005

Recall

Recall Number
Z-1388-05
Event Number
32748
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
LOJ
Status
Terminated
Root Cause
Other
Initiated
July 18, 2005
Posted
August 23, 2005
Terminated
March 7, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

Access Immunoassay Systems Assay Protocol: Dil-AFP.

Reason

Pipetting for dil-Alpha-fetoprotein when performed on the Access, Access 2 and Synchron LX i 725 has a potential to create splashing which may cause erroneous results.

Action

Letters to customers were sent July 18, 2005. Firm will upgrade software later in the year. Customers are advised to not perform the automated dilution protocol but refer to the method in the directions.

Distribution

Nationwide and Canada

Quantity

311 USA 16 Canada