Cold Storage Solution, 2L, labeled as Waters IGL Cold Storage Solution, Catalog No. CSS/2000/US; and, Belzer UW Cold Storage Solution, Catalog No. BUW_002. Intended for the hypothermic flushing and storage of specific organs in preparation for storage, transportation, and transplantation into a recipient.
Recall
- Recall Number
- Z-1387-2018
- Event Number
- 79378
- Firm
- Preservation Solutions, Inc.
- FEI Number
- 1000221028
- Product Code
- KDN
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- February 26, 2018
- Terminated
- September 9, 2022
- Address
- 1099 Proctor Dr, Elkhorn, WI, 53121-2027
Description
Cold Storage Solution, 2L, labeled as Waters IGL Cold Storage Solution, Catalog No. CSS/2000/US; and, Belzer UW Cold Storage Solution, Catalog No. BUW_002. Intended for the hypothermic flushing and storage of specific organs in preparation for storage, transportation, and transplantation into a recipient.
Bags of solution may have a small leak at the base of the injection port where the port is welded into the body of the bag..
An Urgent: Product Recall Notification letters, dated February 26, 2018, were sent to consignees. The instructions include, halting distribution of any 2 Liter bags in inventory; and, arranging for the return of affected units. Preservation Solutions, Inc. provided a notification letter for their consignees to provide to their customers. The letter instructed customers to return any 2 Liter bags to their supplier, and to complete and return the acknowledgement form.
Worldwide Distribution -- USA and Europe.
31,104 units