Blood Gas and Co-Ox2 Electrolyte & Metabolyte Analyzers, ABL-700 Series, manufactured from 8/1998 through 11/2004.
Recall
- Recall Number
- Z-1386-06
- Event Number
- 34963
- Firm
- Radiometer America Inc
- FEI Number
- 1523456
- Product Code
- JFP
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 29, 2005
- Posted
- August 10, 2006
- Terminated
- April 17, 2012
- Address
- 810 Sharon Dr, Westlake, OH, 44145-1598
Description
Blood Gas and Co-Ox2 Electrolyte & Metabolyte Analyzers, ABL-700 Series, manufactured from 8/1998 through 11/2004.
Under certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).
The consignees were notified of this recall by a letter, dated 9/29/2005. The letter informs the customers of this potential problem and the availability of new updated software which will correct the problem. The letter goes on to inform the customer that a Radiometer Service Engineer wil be contacting them shortly in order to arrange for a convenient time to install the software update. Along with the letter, the customer is provided a copy of an additional page and a CD concerning the function and operation of the new software, which is to be added to the customer''s Operator''s Manual.
Nationwide.
1615 devices