Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216) Lot No. 674653

Recall: Z-1385-06 · Initiated July 25, 2006

Recall

Recall Number
Z-1385-06
Event Number
35791
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
GTQ
Status
Terminated
Root Cause
Other
Initiated
July 25, 2006
Posted
August 9, 2006
Terminated
April 18, 2007
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216) Lot No. 674653

Reason

The vial containing the R2 reagent is labeled as R1 reagent and if placed in the R1 slot of the analyzer, erroneous results will be obtained. (extremely low patient and control results)

Action

Consignees were notified via recall letter dated 7/25/06 to cease using and to discard the product.

Distribution

Nationwide including the states of Arizona, California, New Jersey and Pennsylvania.

Quantity

20 kits.