FDA Recall Terminated

SenDx Medical Inc ABL80FLEX Blood Gas, pH, Electrolyte, and Metabolite System The ABL80 FLEX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, and hematocrit in whole blood. The ABL80 FLEX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Recall: Z-1384-2012 · Initiated January 30, 2012

Recall

Recall Number
Z-1384-2012
Event Number
61178
Firm
Sendx Medical Inc
FEI Number
2027541
Product Code
CHL
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 30, 2012
Posted
April 9, 2012
Terminated
July 13, 2012
Address
1945 Palomar Oaks Way, Carlsbad, CA, 92011-1300

Description

SenDx Medical Inc ABL80FLEX Blood Gas, pH, Electrolyte, and Metabolite System The ABL80 FLEX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, and hematocrit in whole blood. The ABL80 FLEX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Reason

The product can cause the analyzer to report values outside the published performance specifications for pH, pCO2, cNa+, cK+, cCa2+ and cCl-.

Action

SenDx Medical sent Customers notification letters dated January 30th, 2012, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer. We have recently become aware of a possible condition in a limited number of ABLBO FLEX analyzers configured with a specific Control PCe lerinted Circuit Board), which can cause the analyzer to report values outside the published performance specifications for pH, pCOz, cNa*, cK*, cCa2* and cCl-. Customers were instructed to immediately implement Alternative 1 or 2 until the signal data PCB is replaced. Alterative I: Stop the use of all affected analyzers until corrective actions can be performed on these analyzers. Alternative 2: Verify proper function of the affected analyzers immediately following every patient sample measurement by measuring a manual QC sample. Customers were asked to complete the fax form acknowledging receipt of the notice and provided a contact for Technical Support at 1-800-736-0600 opt.4.

Distribution

One US consignee in state of OH. Other consignees are international such as Australia, France, Netherlands, Denmark, China, United Kingdom, Turkey, Switzerland, Japan, India, Poland, Spain, and Germany. Distribution list submitted with supporting doc.s

Quantity

819 analyzers and 44 signal data PCBs