FDA Recall Terminated

AUTOMIX Plus High Speed Compounder System, product 2M8075; automated nutrition compounders; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 The Automix Compounder Systems are electromechanical positive displacement fluid compounding devices used in the admixture of large volume parenteral nutrition formulations via gravimetric weighing.

Recall: Z-1384-2011 · Initiated January 27, 2011

Recall

Recall Number
Z-1384-2011
Event Number
57790
Firm
Baxter Healthcare Corp. Rt.
FEI Number
1416980
Product Code
LHI
Status
Terminated
Root Cause
Other
Initiated
January 27, 2011
Posted
February 22, 2011
Terminated
February 16, 2012
Address
120 & Wilson Rd, Round Lake, IL, 60073

Description

AUTOMIX Plus High Speed Compounder System, product 2M8075; automated nutrition compounders; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 The Automix Compounder Systems are electromechanical positive displacement fluid compounding devices used in the admixture of large volume parenteral nutrition formulations via gravimetric weighing.

Reason

The AUTOMIX compounder may not respond correctly to the key pressed by the operator using the keypad. The incorrect keypad response issue can occur both in manual data entry mode (STD Mode) and when entering order numbers to download from host software (MAN ID Mode). It is possible that an incorrect volume and/or an incorrect solution could be admixed into the final bag.

Action

Baxter Healthcare Corp. sent Urgent Device Correction letters dated January 27, 2011, via first class mail to all affected customers. informing personnel that use Automix compounders to exercise care when entering data, and verify the accuracy of the keypad response. The accounts were requested to convey this information to all users of the Automix compounders, emphasizing the need to verify data accuracy each time the keypad is used. The accounts were also requested to complete the attached customer reply form listing the model and serial number of each of the Automix compounders at the facility and confirming their receipt of the letter, and fax it to Baxter at the number provided on the form. Any questions regarding the communication were directed to their local Baxter representative, and technical questions were directed to Baxter's Global Technical Services at 1-800-626-2667. After determining the root cause of the incorrect keypad responses, Baxter issued a follow-up correction letter on 4/27/11 outlining the recommended work practices to reduce fluid penetration into the Automix compounder keypad/control module.

Distribution

Worldwide Distribution - USA - Nationwide including Puerto Rico, and internationally to Brazil, Canada, Colombia, Costa Rica, Czech Republic, El Salvador, Egypt, Germany, Guatemala, Honduras, Luxembourg, Mexico, Panama, Saudi Arabia, Singapore, Switzerland, Taiwan, Trinidad, Turkey, United Arab Emirates and United Kingdom.

Quantity

840 units