FDA Recall Open, Classified

DigitalDiagnost C50 1.1-intended for use in generating radiographic images of human anatomy Model Number : 712204

Recall: Z-1383-2023 · Initiated March 9, 2023

Recall

Recall Number
Z-1383-2023
Event Number
91915
Firm
Philips North America Llc
FEI Number
3016618143
Product Code
KPR
Status
Open, Classified
Root Cause
Equipment maintenance
Initiated
March 9, 2023
Posted
April 12, 2023
Address
222 Jacobs St, Cambridge, MA, 02141-2289

Description

DigitalDiagnost C50 1.1-intended for use in generating radiographic images of human anatomy Model Number : 712204

Reason

If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users

Action

Philips issued Urgent Medical Device Correction Letter to US Consignees on 09-Mar-2023. Letter states reason for recall, health risk and action to take: Follow the Planned Maintenance Program in the Addendum to the Instructions for Use in Appendix A of this notice for the safe use of your system(s). " If the Ceiling Suspension is not able to be kept in its current position by itself and descends after the power to the system is off, contact your local Philips representative to perform the maintenance of the telescopic carriage. " Place this URGENT Medical Device Correction Notice with the documentation of your system(s). " Complete the enclosed Customer Reply Form and send it back to confirm that users have reviewed and understood this URGENT Medical Device Correction Notice. Philips is informing customers through this URGENT Medical Device Correction Notice and providing the Addendum to the Instructions for Use of the DigitalDiagnost C50 system. If you need additional information or support concerning this issue, please contact the Customer Care -9377).

Distribution

Nationwide including Puerto Rico and Virgin Islands Foreign: ARGENTINA Australia Austria Bosnia and Herzegovina Bulgaria China Colombia Cyprus Czech Republic Egypt France French Polynesia Germany Greece Hungary Indonesia Italy Jordan Kenya Latvia Lithuania Macedonia Malta Mexico Netherlands New Zealand Philippines Poland Portugal Puerto Rico Russian Federation Saudi Arabia Serbia Slovenia South Africa Spain Sri Lanka Switzerland Tanzania Thailand United Arab Emirates Viet Nam Yemen

Quantity

10 units US; 283 units OUS