FDA Recall Terminated

XenMatrix Surgical Graft ,Sterile, Porcine Dermal Matrix 19 x 35 cm Rectangle Catalog #: 1161935

Recall: Z-1382-2011 · Initiated January 7, 2011

Recall

Recall Number
Z-1382-2011
Event Number
57566
Firm
Davol, Inc., Sub. C. R. Bard, Inc.
FEI Number
1213643
Product Code
FTM
Status
Terminated
Root Cause
Process control
Initiated
January 7, 2011
Posted
March 2, 2011
Terminated
October 31, 2013
Address
100 Crossings Blvd, Warwick, RI, 02886-2850

Description

XenMatrix Surgical Graft ,Sterile, Porcine Dermal Matrix 19 x 35 cm Rectangle Catalog #: 1161935

Reason

Davol cannot guarantee that all Xenomatic product released from July to October 2010 meets FDA guidelines for endotoxin limits; and therefore, the grafts are potentially contaminated with bacteria.

Action

Davol issued recall notification to accounts by letter dated 1/06/11 requesting recalled units be returned. contact Davol Customer Service at 1-800-556-6275 for instructions on how to return product to Davol. A fax sheet attached to the letter to be completed and fax to 1-401-825-8753.

Distribution

Nationwide

Quantity

886 units