Ivenix Infusion System (IIS)
Recall
- Recall Number
- Z-1381-2022
- Event Number
- 90314
- Firm
- Ivenix, Inc.
- FEI Number
- 3014732157
- Product Code
- PHC
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- April 22, 2022
- Terminated
- May 20, 2024
- Address
- 50 High St, Ste 50, North Andover, MA, 01845-2620
Description
Ivenix Infusion System (IIS)
Downstream occlusion alarm is sometimes immediately followed by a pump problem alarm.
On April 22, 2022, the firm notified affected customers of the product issue through a letter titled "Important Safety Update." The firm is working on a software update to correct the issue. It will be available in approximately 60 days. The firm will inform customers when the update is ready to install. As interim corrections, the firm recommends that any pump that is infusing into the downstream access port have an infusion pressure less than 525 mm Hg. Alternatively, the Occlusion Retry Attempts in the Infusate Library Care Profile Pump Setting should be set to 0. Care profile pump settings are described in the Infusate Library and Pump Management Instructions for Use. If you have additional questions, please contact Customer Support: [email protected] or 1-978-775-8100
US Nationwide distribution in the states of NJ and WI.
1335 devices