FDA Recall Terminated

Minerva-Combi Patient Lift, model ML-30; this lift comes with interchangable passive and active lifting units; Manufactured by Medibo-Technics, Hasseltsebaan 24, 3940 Hechtel-Eksel, Belgium and distributed by MasterCare Patient Equipment, Inc., Hwy 61 West Industrial Park, Columbus, NE 68601

Recall: Z-1381-04 · Initiated August 4, 2004

Recall

Recall Number
Z-1381-04
Event Number
29662
Firm
Medibo N.V. Heikant 5 Hanmont, Achel Belgium
FEI Number
3004468271
Product Code
FSA
Status
Terminated
Root Cause
Other
Initiated
August 4, 2004
Posted
August 26, 2004
Terminated
December 13, 2004

Description

Minerva-Combi Patient Lift, model ML-30; this lift comes with interchangable passive and active lifting units; Manufactured by Medibo-Technics, Hasseltsebaan 24, 3940 Hechtel-Eksel, Belgium and distributed by MasterCare Patient Equipment, Inc., Hwy 61 West Industrial Park, Columbus, NE 68601

Reason

The hanger bar may detach from the jib yoke, the M8 bolt in the foot pedal assembly may come loose, and some units may have faulty actuator brackets on the mast assembly, resulting in the patient lift becoming unstable and allowing the patient to fall.

Action

Customer Advisory Notice (MB/0407/CAN-001) letters dated 8/6/04 were sent by FedEx Ground Service on 8/4/04 to the end users, informing them of the possibility of the hanger bar assembly to detach from the jib due wear of the pivot assembly within the yoke. They were also advised that the bolt in the foot pedal assembly could come loose, rendering the lift unstable, and that some of the lifts may have faulty actuator brackets on the mast assembly which could result in the collapse of the hoist. The users were requested to inspect their lifts or take them out of service pending service by an authorized service engineer. The users were requested to complete customer response form indicating the model numbers and serial numbers of the lifts at the facility, both in use and out of service, and fax the form back to the Minerva Recall Department at 1-888-594-2756 within 5 days. Within two weeks an authorized service engineer will contact each facility to schedule an appointment to service the lifts at Medibo''s expense. Any questions were directed to the Minerva Recall Department, at 1-888-402-6448.

Distribution

California, Florida, Louisiana, Massachusetts, New Hampshire, New Jersey, New York, Ohio, Oregon and Pennsylvania.

Quantity

15 units