FDA Recall Terminated

CDI Blood Parameter Monitoring System 500, 500AHCT. Intended for use during cardiopulmonary bypass procedures.

Recall: Z-1379-2012 · Initiated February 3, 2012

Recall

Recall Number
Z-1379-2012
Event Number
60152
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DRY
Status
Terminated
Root Cause
Labeling design
Initiated
February 3, 2012
Posted
April 6, 2012
Terminated
August 29, 2013
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

CDI Blood Parameter Monitoring System 500, 500AHCT. Intended for use during cardiopulmonary bypass procedures.

Reason

Replacement of the SBC batteries (system batteries) should be managed as "routine maintenance" instead of as an in-house Service Procedure because some systems may be relying on batteries beyond their useful life.

Action

On 2/3/12 all consignees received an URGENT MEDICAL DEVICE RECALL letter via Federal Express. The letter described the reason for the recall, identified the affected product population, what to do in the event of a failure, the potential hazard, and correction. Terumo CVS will update the Operator's Manual for the CDI 500 with instructions to replace the battery within its useful life. Users are to review the notice and ensure that all users are made aware of the issue. In addition, the Customer Response Form should be completed and returned. Users can contact their local Terumo Cardiovascular Systems representative or call 1-800-521-2818 with questions or concerns, to check the last time system batteries were replaced, or to schedule maintenance for systems approaching 10 years.

Distribution

Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, WV, WA, and WI and the countries of Japan, Germany, Barbados, Saudi Arabia, Columbia, Uruguay, Ecuador, Trinidad, Costa Rica, Paraguay, Panama, Belgium, Jordan, Libya, Mexico, Brazil, Honduras, Canada, Korea, Dominica Republic, El Salvador, Guatemala, Vietnam, Chile, Hong Kong, United Arab Emirates, Singapore, Australia, Malaysia, Taiwan, Philippines, Mexico, Thailand, Vietnam, and South Africa.

Quantity

3,829 units