FDA Recall Terminated

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.

Recall: Z-1378-2018 · Initiated January 12, 2018

Recall

Recall Number
Z-1378-2018
Event Number
79711
Firm
Cook Inc.
FEI Number
1820334
Product Code
MQG
Status
Terminated
Root Cause
Device Design
Initiated
January 12, 2018
Posted
April 13, 2018
Terminated
May 8, 2020
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.

Reason

The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.

Action

Cook plans to begin notifying consignees of the recall on January 12, 2018 Written recall communication letters will be mailed via courier to US consignees

Distribution

Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)

Quantity

4,451