FDA Recall Terminated

Spetzler Claw Tip, Universal - 5450-800-311 Spetzler Open Angle Micro Claw Tip, Universal - 5450-800-313 The Spetzler Claw" Tips are bone cutting tips which are a part of Strykers Neuro Spine Ear, Nose and Throat (ENT) Sonopet Portfolio. They are removable attachments intended to be used with the Sonopet Ultrasonic Handpiece.

Recall: Z-1378-2016 · Initiated March 22, 2016

Recall

Recall Number
Z-1378-2016
Event Number
73610
Firm
Stryker Instruments Div. of Stryker Corporation
FEI Number
1811755
Product Code
LFL
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
March 22, 2016
Posted
April 13, 2016
Terminated
July 19, 2016
Address
4100 E Milham Ave, Portage, MI, 49002-9704

Description

Spetzler Claw Tip, Universal - 5450-800-311 Spetzler Open Angle Micro Claw Tip, Universal - 5450-800-313 The Spetzler Claw" Tips are bone cutting tips which are a part of Strykers Neuro Spine Ear, Nose and Throat (ENT) Sonopet Portfolio. They are removable attachments intended to be used with the Sonopet Ultrasonic Handpiece.

Reason

The Spetzler Claw Tip was packaged with the incorrect Tip Sleeve Assembly. The Tip Sleeve Assembly included in the packaging is the incorrect size. Risk to Health: If the Tip Sleeve Assembly does not fit the Spetzler Claw" Tip, a surgical delay may occur while an alternate Tip Sleeve Assembly is retrieved.

Action

On 3/22/2016, URGENT MEDICAL DEVICE RECALL NOTIFICATION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments: Kara Spath 269-389-4518 [email protected]

Distribution

Distributed in the states of AZ, CA, CO, CT, IL, IN, MD, MI, MN, NY, PA, SC, TN, WA, WI, and VA, and in the NETHERLANDS.

Quantity

140 units