ADVIA 2120 systems -Automated Complete Blood Cell and Differential Cell Counter
Recall
- Recall Number
- Z-1376-06
- Event Number
- 35893
- Firm
- Bayer Healthcare, LLC (Diagnostics Division)
- FEI Number
- 2432235
- Product Code
- GKL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 16, 2006
- Posted
- August 4, 2006
- Terminated
- May 24, 2012
- Address
- 511 Benedict Avenue, Tarrytown, NY, 10591-5097
Description
ADVIA 2120 systems -Automated Complete Blood Cell and Differential Cell Counter
The ADVIA 2120 has reported ; highly intermittent low results on all primary results: White Blood Cells (WBC), Red Blood Cells (RBC), Hemoglobin (HGB) and Platelets (PLT) and an abnormal baso cytogram.
Bayer HealthCare LLC issued Support Bulletins to all affected Bayer HealthCare LLC Branches on 3/16/06 in order to inform them of the issue and to provide them with instructions for this ''field correction''. Bayer also issued Customer Bulletins to the Branches both in the United States and elsewhere for communication with the affected customers.
Worldwide distribution-Domestic consignees consist of hospitals and clinical laboratories. Foreign countries (units): Canada (1), China (4), Singapore (1), Malaysia (2), Thailand (1), Taiwan (1), Vietnam (1), Australia (3), Korea (25), Japan (25), Israel (4), Saudi Arabia (4), Russia (1), Turkey (1), United Arab Emirates (1), South Africa (2), United Kingdom (31), Germany (1), Poland (9), Norway (1), Spain (48), Finland (7), Denmark (4), Portugal (14), Italy (5), Netherlands (3), France (46), Ireland (4), India (7), Austria (1).
60 units Domestic, 258 units Foreign