FDA Recall Terminated

ADVIA 2120 systems -Automated Complete Blood Cell and Differential Cell Counter

Recall: Z-1376-06 · Initiated March 16, 2006

Recall

Recall Number
Z-1376-06
Event Number
35893
Firm
Bayer Healthcare, LLC (Diagnostics Division)
FEI Number
2432235
Product Code
GKL
Status
Terminated
Root Cause
Other
Initiated
March 16, 2006
Posted
August 4, 2006
Terminated
May 24, 2012
Address
511 Benedict Avenue, Tarrytown, NY, 10591-5097

Description

ADVIA 2120 systems -Automated Complete Blood Cell and Differential Cell Counter

Reason

The ADVIA 2120 has reported ; highly intermittent low results on all primary results: White Blood Cells (WBC), Red Blood Cells (RBC), Hemoglobin (HGB) and Platelets (PLT) and an abnormal baso cytogram.

Action

Bayer HealthCare LLC issued Support Bulletins to all affected Bayer HealthCare LLC Branches on 3/16/06 in order to inform them of the issue and to provide them with instructions for this ''field correction''. Bayer also issued Customer Bulletins to the Branches both in the United States and elsewhere for communication with the affected customers.

Distribution

Worldwide distribution-Domestic consignees consist of hospitals and clinical laboratories. Foreign countries (units): Canada (1), China (4), Singapore (1), Malaysia (2), Thailand (1), Taiwan (1), Vietnam (1), Australia (3), Korea (25), Japan (25), Israel (4), Saudi Arabia (4), Russia (1), Turkey (1), United Arab Emirates (1), South Africa (2), United Kingdom (31), Germany (1), Poland (9), Norway (1), Spain (48), Finland (7), Denmark (4), Portugal (14), Italy (5), Netherlands (3), France (46), Ireland (4), India (7), Austria (1).

Quantity

60 units Domestic, 258 units Foreign