FDA Recall Terminated

ProSpecT Cryptosporidium Microplate Assay. Catalog #2454096.

Recall: Z-1376-04 · Initiated May 9, 2002

Recall

Recall Number
Z-1376-04
Event Number
29681
Firm
Remel, Inc.
FEI Number
1924669
Product Code
MHJ
Status
Terminated
Root Cause
Other
Initiated
May 9, 2002
Posted
August 17, 2004
Terminated
September 15, 2004
Address
12076 Santa Fe Trail Dr (PO Box 14478, zip 66285-4478), Lenexa, KS, 66215-3519

Description

ProSpecT Cryptosporidium Microplate Assay. Catalog #2454096.

Reason

A potential stability problem has been identified with the Positive Controls in the referenced kit lots of the ProSpecT Cryptosporidium Microplate Assay. The Positive Control may NOT perform as intended through its expiration date and may invalidate an otherwise acceptable run. Quality control testing indicates the other reagents in these kits are performing as intended.

Action

A potential stability problem has been identified with the positive control in the ProSpecT CryptoSporidium Microplate Assay. The positive control may not perform as intended through its expiration date and may invalidate an otherwise acceptable run.

Distribution

in the US

Quantity

388 kits (187 - 910717, 201 - 912101)