AMO brand WaveScan WaveFront System, OPM software versions 3.67 and 3.67, Part Number: 0070-1531, Catalog Numbers: 0070-1664,0070-1649,0070-1654: Version 3.67 0070-1721: Version 3.671 0070-1662: Version 3.67, US version; 0070-1668; IVersion 3.67 International version Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The WaveScan WaveFront System is a class III accessory device used in conjunction with the VISX STAR S4 and STAR S4 IRTM Excimer Laser Systems, which are indicated for wavefront-guided laser assisted in situ keratomileusis (LASIK) eye surgery in certain patient populations. The WaveScan WaveFront System is a diagnostic instrument intended for the automated measurement, analysis, and recording of refractive errors of the eye. The WaveScan System software uses the diagnostic data of the eye's refractive errors and wavefront aberrations to calculate the target treatment shape and then generates the commands for the VISX STAR laser to create the target shape on the cornea.
Recall
- Recall Number
- Z-1374-2011
- Event Number
- 56394
- Firm
- AMO Manufacturing USA, LLC
- FEI Number
- 3006695864
- Product Code
- LZS
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- July 13, 2010
- Posted
- February 18, 2011
- Terminated
- June 22, 2012
- Address
- 510 Cottonwood Dr, Milpitas, CA, 95035-7403
Description
AMO brand WaveScan WaveFront System, OPM software versions 3.67 and 3.67, Part Number: 0070-1531, Catalog Numbers: 0070-1664,0070-1649,0070-1654: Version 3.67 0070-1721: Version 3.671 0070-1662: Version 3.67, US version; 0070-1668; IVersion 3.67 International version Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The WaveScan WaveFront System is a class III accessory device used in conjunction with the VISX STAR S4 and STAR S4 IRTM Excimer Laser Systems, which are indicated for wavefront-guided laser assisted in situ keratomileusis (LASIK) eye surgery in certain patient populations. The WaveScan WaveFront System is a diagnostic instrument intended for the automated measurement, analysis, and recording of refractive errors of the eye. The WaveScan System software uses the diagnostic data of the eye's refractive errors and wavefront aberrations to calculate the target treatment shape and then generates the commands for the VISX STAR laser to create the target shape on the cornea.
The failure to comply with Minimum System Requirements, increased system errors could result, and a miscreated treatment table (incorrect LASIK refractive treatment) could result.
Abbott Medical Optics, Inc. sent an Urgent Medical Device Field Correction letter to all affected customers starting July 15, 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to take the following actions: Surgeons were instructed to complete the facsimile attachment to acknowledge receipt of the notification letter and fax to 714-247-4530. For those using the affected product a 3.68 OPM disc must be requested by e-mailing a request to [email protected]. For any questions please contact AMO Technical Service Representatiave at 1-877-266-4543.
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Great Britian, Greece, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Malaysia, Mexico, Netherland, New Zealand, Peru, Philippines, Portugal, Russia, Russia Federation, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Taiwan, ThailandChina, India, South Korea, Singapore, Taiwan, Thailand, Turkey, United Arab Emirates, Venezuela, and Vietnam.