FDA Recall Terminated

Lifeshield, latex-free, Primary I.V. Set, convertible pin, 100 inch with OPTION-LOK and VENOSET, 15 drops/mL, a sterile Rx single use device, Hospira Inc, Lake Forest, IL USA. Made in Costa Rica, list 12574-48.

Recall: Z-1374-2010 · Initiated November 24, 2009

Recall

Recall Number
Z-1374-2010
Event Number
54079
Firm
Hospira Inc.
FEI Number
3013319212
Product Code
FPA
Status
Terminated
Root Cause
Component design/selection
Initiated
November 24, 2009
Posted
April 15, 2010
Terminated
January 29, 2012
Address
375 N Field Dr, Lake Forest, IL, 60045-2513

Description

Lifeshield, latex-free, Primary I.V. Set, convertible pin, 100 inch with OPTION-LOK and VENOSET, 15 drops/mL, a sterile Rx single use device, Hospira Inc, Lake Forest, IL USA. Made in Costa Rica, list 12574-48.

Reason

Reports of difficulty in adjusting the flow rate, which could result in flow rate variances.

Action

Hospira sent the Urgent Device Recall letters dated 11/25/09 to their direct accounts of the affected products and requested these consignees to notify their customers of the recall. The accounts were informed that Hospira ceased manufacturing the I.V. gravity and pump administration sets with the CAIR clamp manufactured by Borla S.P.A. of Italy in March 2008, due to customer reports related to a difference in the zone of control used to adjust the flow compared to the original CAIR clamp design. To avoid the possibility of flow rate variances, Hospira is recalling all remaining sets with the Borla designed CAIR clamp. The accounts were requested to quarantine all inventory of the affected lots, complete the enclosed reply form indicating the amount of product being returned, fax a copy of the completed form to Stericycle at 1-877-377-5114, and include a copy with the returned goods to Stericycle at 2670 Executive Drive, Suite A, Indianapolis, IN 46241.

Distribution

Worldwide Distribution -- U.S., Puerto Rico, Dominican Republic, Bahamas, Virgin Islands, Argentina, Australia, Bahrain, Belgium, Canada, China, Denmark, Ecuador, EI Salvador, Finland, France, Germany, Greece, Guatemala, Hong Kong, Ireland, Italy, Jordan, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uruguay, and United Arab Emirates.

Quantity

237,504 sets