FDA Recall Terminated

CHOL+HDL+GLU test strips, REF 2412

Recall: Z-1373-2019 · Initiated January 16, 2019

Recall

Recall Number
Z-1373-2019
Event Number
82331
Firm
Polymer Technology Systems, Inc.
FEI Number
3015525148
Product Code
CGA
Status
Terminated
Root Cause
Device Design
Initiated
January 16, 2019
Terminated
June 24, 2020
Address
7736 Zionsville Rd, Indianapolis, IN, 46268-2175

Description

CHOL+HDL+GLU test strips, REF 2412

Reason

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed

Action

The firm initially notified some of their customers from January through March 2019 with a Customer Bulletin. On March 28, 2019, the firm issued an updated Urgent Medical Device Recall letter to all affected customers. The letter alerted customers to the potential for test strips to slip out of the analyzer. Actions to be taken by the Customer/User: 1. Closely follow the instructions for proper testing. " Make sure the analyzer is sitting on a flat, stable surface. Do not perform testing with the analyzer held in the hand. " After inserting the test strip(s), do not move the analyzer until testing is complete and the test strip is removed. " As stated in the instructions for use, if an unexpected test result is given, the test should be repeated. There is no need to return or destroy the product. 2. Distributors should immediately forward this notice to all customers/users who may have received this product. 3. Acknowledge receipt of this instruction by completing the attached form, even if you do not have affected product on hand. If you have any questions, please contact Karmen Mercer, Director of Global Customer Care directly, Monday through Friday, 8:00 AM to 5:00 PM, Eastern Time at +1 317-870-5610 or (U.S. toll-free) 877-570-8610.

Distribution

Domestic distribution nationwide.

Quantity

67643 total