FDA Recall Open, Classified

VITROS Performance Verifier I

Recall: Z-1372-2024 · Initiated February 27, 2024

Recall

Recall Number
Z-1372-2024
Event Number
94174
Firm
Ortho-Clinical Diagnostics
FEI Number
1000305841
Product Code
JJY
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
February 27, 2024
Posted
March 22, 2024
Address
1000 Lee Rd, Rochester, NY, 14606-4250

Description

VITROS Performance Verifier I

Reason

The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.

Action

Consignees were sent an URGENT PRODUCT CORRECTION NOTIFICATION dated 2/27/24 informing them that for product Lot Q1174, diluent Lot M9815 was found in some sales units instead of the expected diluent Lot Q1175, and for product Lot R1176, diluent Lot P9967 was found in some sales units instead of the expected diluent Lot R1177. Consignees were asked to inspect their inventory to see if any mismatched product from affected lots was present and to discard any affected units. The provided Confirmation of Receipt form is to be returned by 3/8/24. Results received using affected devices are to be reviewed for any unexpected QC shift. The recall notification is to be forwarded if any devices were further distributed and the notification maintained in the User Documentation. Customers with any questions can call Global Services at 1-800-421-3311.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Quantity

10,911 units