FDA Recall Open, Classified

Aptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent, Enzyme Reagent, Promoter Reagent, Amplification Reconstitution Solution, Enzyme Reconstitution Solution, Promoter Reconstruction solution, Target capture Reagent, Reconstitution Collars.

Recall: Z-1371-2024 · Initiated September 14, 2023

Recall

Recall Number
Z-1371-2024
Event Number
94038
Firm
Hologic, Inc.
FEI Number
2024800
Product Code
PAB
Status
Open, Classified
Root Cause
Use error
Initiated
September 14, 2023
Posted
March 22, 2024
Address
10210 Genetic Center Dr, San Diego, CA, 92121-4362

Description

Aptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent, Enzyme Reagent, Promoter Reagent, Amplification Reconstitution Solution, Enzyme Reconstitution Solution, Promoter Reconstruction solution, Target capture Reagent, Reconstitution Collars.

Reason

Use of assay along with plasma specimen may result in error code which would invalidate sample.

Action

On February 23, 2024, Hologic issued a "Urgent Medical Device Recall Update" via FedEx. The update notified consignees of a sample dilution workflow for specific plasma specimens. On September 14, 2023 Hologic issued a "Urgent Medical Device Recall Notification" to affected consignees, via Fed Ex. Hologic provided the following information to customers: 1. Continue to use the Aptima CMV Quant assay. 2.When obtaining an ML2 error, please run the Mag Wash Clean Procedure and Contact Hologic Technical Support 3. With the Aptima CMV Quant assay, it is not recommended to retest a sample that was previously invalidated due to an ML2 error on the Panther instrument. 4. Complete Customer Response Form On (date) the firm issued a Customer Technical Bulletin, to consignees describing a sample dilution workflow for plasma specimens.

Distribution

Worldwide - US Nationwide distribution in the states of CA, CT, FL, ID, IL, MD, MI, MO, TN, TX, WI and the countries of Australia, Germany, Denmark, Spain, France, Great Britain, Italy, Norway, Qatar, Sweden.

Quantity

684 kits