FDA Recall
Terminated
ACCU-CHEK Voicemate System, Advantage Blood Glucose Meter and Voice Unit - Spanish language version for the blind or visually impaired, Cat. no. 3040208. Ref. no. 03040208001 with meter model no. 768.
Recall: Z-1371-06
·
Initiated June 22, 2006
Recall
- Recall Number
- Z-1371-06
- Event Number
- 35022
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- LFR
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 22, 2006
- Posted
- August 5, 2006
- Terminated
- April 7, 2008
- Address
- 9115 Hague Rd, Indianapolis, IN, 46250-0416
Description
ACCU-CHEK Voicemate System, Advantage Blood Glucose Meter and Voice Unit - Spanish language version for the blind or visually impaired, Cat. no. 3040208. Ref. no. 03040208001 with meter model no. 768.
Reason
The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Action
A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Distribution
Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
Quantity
55,189 meters (for all Recalls Z-1354-1371-06)