FDA Recall Open, Classified

Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.

Recall: Z-1370-2026 · Initiated December 31, 2025

Recall

Recall Number
Z-1370-2026
Event Number
98315
Firm
Medica Corporation
FEI Number
1220972
Product Code
CEM
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
December 31, 2025
Posted
February 13, 2026
Address
5 Oak Park Dr, Bedford, MA, 01730-1430

Description

Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.

Reason

A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.

Action

An URGENT FIELD SAFETY NOTICE was sent to consignees on 01/09/25 informing them that potassium results obtained using affected devices in capillary mode may be falsely elevated. The FSN instructed consignees not to report potassium results obtained in capillary mode using affected lots on the Easy Stat Analyzer. The FSN is to be forwarded to end users and to where product was further distributed. Consignees are to contact their local representative with any questions.

Distribution

International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.

Quantity

1,578 units