BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
Recall
- Recall Number
- Z-1370-2015
- Event Number
- 70678
- Firm
- BD Biosciences, Systems & Reagents
- FEI Number
- 2916837
- Product Code
- OYE
- Status
- Terminated
- Root Cause
- Mixed-up of materials/components
- Initiated
- March 6, 2015
- Posted
- June 26, 2015
- Terminated
- April 4, 2016
- Address
- 2350 Qume Dr, San Jose, CA, 95131-1812
Description
BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.
BD Biosciences sent an Important Product Recall letter on March 6, 2015 by certified mail and Field Change Notification (FCN FCB15-01-660312 directing Field Service Engineers to correct the circuit breaker is in process. Customers with questions were instruct to contact BD Customer Support at 1-877-232-8995 (prompt 2) in the United States. For customers outside the US were instructed to contact their local BD Biosciences representative or distributor. For questions regarding this recall call 408-954-6080.
Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim
72 devices