FDA Recall Terminated

BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.

Recall: Z-1370-2015 · Initiated March 6, 2015

Recall

Recall Number
Z-1370-2015
Event Number
70678
Firm
BD Biosciences, Systems & Reagents
FEI Number
2916837
Product Code
OYE
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
March 6, 2015
Posted
June 26, 2015
Terminated
April 4, 2016
Address
2350 Qume Dr, San Jose, CA, 95131-1812

Description

BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.

Reason

The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.

Action

BD Biosciences sent an Important Product Recall letter on March 6, 2015 by certified mail and Field Change Notification (FCN FCB15-01-660312 directing Field Service Engineers to correct the circuit breaker is in process. Customers with questions were instruct to contact BD Customer Support at 1-877-232-8995 (prompt 2) in the United States. For customers outside the US were instructed to contact their local BD Biosciences representative or distributor. For questions regarding this recall call 408-954-6080.

Distribution

Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim

Quantity

72 devices