FDA Recall
Terminated
ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter.
Recall: Z-1370-05
·
Initiated June 7, 2005
Recall
- Recall Number
- Z-1370-05
- Event Number
- 32100
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- NBW
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 7, 2005
- Posted
- August 13, 2005
- Terminated
- July 2, 2007
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter.
Reason
The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards, and held upwards, while the meter is waiting for the sample.
Action
Recall letters dated 6/7/05 were sent to all customers. Distributors were requested to send a copy of the notification to all customers of test strips used with the product. Users are instructed to never operate the meter without the optics cover in place, and this information will be included in test strip boxes.
Distribution
Nationwide.
Quantity
2490