FDA Recall Terminated

EvoTech Endoscope Cleaner & Reprocessor, Product Code 50004 Indicated for use with high level disinfectant CIDEX OPA Concentrate and an enzymatic detergent, CIDEZYME GI, to achieve cleaning and high level disinfection of heat sensitive semi-critical endoscopes.

Recall: Z-1368-2008 · Initiated February 13, 2008

Recall

Recall Number
Z-1368-2008
Event Number
46906
Firm
Advanced Sterilization Products
FEI Number
3003662624
Product Code
FEB
Status
Terminated
Root Cause
Other
Initiated
February 13, 2008
Posted
August 8, 2008
Terminated
November 29, 2011
Address
33 Technology Dr, Irvine, CA, 92618-2346

Description

EvoTech Endoscope Cleaner & Reprocessor, Product Code 50004 Indicated for use with high level disinfectant CIDEX OPA Concentrate and an enzymatic detergent, CIDEZYME GI, to achieve cleaning and high level disinfection of heat sensitive semi-critical endoscopes.

Reason

Cleaning cycles are being cancelled. Specific issues are: a) Water heater tank is rusting resulting in the in-line 0.2 micron filter prematurely clogging, b) Interface board malfunction resulting in a premature lid opening, and c) Minor fluid leaks within the system (commonly at the 0.2 micron filter attachment points).

Action

Feburary 13, 2008 ASP issued an Urgent Product Recall Letter. ASP recommends that customers use an alternative reprocessing method until they return the system. However, if they do not have an alternate method for processing endoscopes and choose to continue using the EvoTech ECR System during this time, endoscopes processed through completed cycles in accordance with the User's Guide will be appropriately cleaned and highlevel disinfected. ASP is prepared to tailor an alternative solution to minimize the interruption to the customer's processing needs.

Distribution

Worldwide Distribution --- including USA and countries of Canada and Japan.

Quantity

64 units