FDA Recall Open, Classified

Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems. Product Code: 6199960

Recall: Z-1367-2022 · Initiated May 26, 2022

Recall

Recall Number
Z-1367-2022
Event Number
90408
Firm
Ortho-Clinical Diagnostics, Inc.
FEI Number
1000136573
Product Code
QKO
Status
Open, Classified
Root Cause
Incorrect or no expiration date
Initiated
May 26, 2022
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems. Product Code: 6199960

Reason

On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).

Action

Ortho Clinical Diagnostics issued Important Correction Notification letter (CL2022-135) on May 26, 2022. Letter states reason for recall, health risk and action to take: When using VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack, do not store on-analyzer for longer than 6 weeks. " Until a lot update occurs, OAS for VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack will temporarily require manual monitoring and removal of reagent packs from analyzer no longer than 6 weeks after packs have been opened and loaded onto the analyzer. " Complete the enclosed Confirmation of Receipt form no later than June 9, 2022. " Post this notification by each VITROS ECi/ECiQ/3600/5600/XT 7600 System in your laboratory until further notice. " Please forward this notification if VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack was distributed outside of your facility. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311

Distribution

Worldwide distribution - US Nationwide and the countries of Colombia, India, Singapore.

Quantity

20858