FDA Recall Terminated

AccuData GTS Plus AccuData GTS For Blood Glucose Monitoring and Automated Data Management, contents Accu-Chek Advantage repalcement Meter and module, cat. no. 454; Ref. no. 03144798001 contains meter model no. 777.

Recall: Z-1367-06 · Initiated June 22, 2006

Recall

Recall Number
Z-1367-06
Event Number
35022
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
CGA
Status
Terminated
Root Cause
Other
Initiated
June 22, 2006
Posted
August 5, 2006
Terminated
April 7, 2008
Address
9115 Hague Rd, Indianapolis, IN, 46250-0416

Description

AccuData GTS Plus AccuData GTS For Blood Glucose Monitoring and Automated Data Management, contents Accu-Chek Advantage repalcement Meter and module, cat. no. 454; Ref. no. 03144798001 contains meter model no. 777.

Reason

The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.

Action

A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.

Distribution

Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.

Quantity

55,189 meters (for all Recalls Z-1354-1371-06)