Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system

Recall: Z-1365-2022 · Initiated May 20, 2022

Recall

Recall Number
Z-1365-2022
Event Number
90423
Firm
Draeger Medical Systems, Inc.
FEI Number
2510954
Product Code
FPL
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 20, 2022
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system

Reason

The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.

Action

On May 20, 2022, the firm began to send Important Field Safety Notice letters to affected customers. Customers were instructed to avoid using the ventilator portion of the system due to the potential for enriched ambient air rising beyond safe levels until a replacement part can be installed and tested. The firm will send customers a replacement kit and testing which will be completed at no charge to the customer. Draeger Service Representatives will call customers and schedule a service visit to update affected systems.