ACCU-CHEK HQ for Blood Glucose Monitoring and Automated Data Management and Communication; Contents include two Accu-Chek Advantage monitors including Cat. no. 2138018 and Cat. no. 2138026. Ref. nos. 12138018001 and 12218542001 both include include meters with model number 777.
Recall
- Recall Number
- Z-1365-06
- Event Number
- 35022
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- LFR
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 22, 2006
- Posted
- August 5, 2006
- Terminated
- April 7, 2008
- Address
- 9115 Hague Rd, Indianapolis, IN, 46250-0416
Description
ACCU-CHEK HQ for Blood Glucose Monitoring and Automated Data Management and Communication; Contents include two Accu-Chek Advantage monitors including Cat. no. 2138018 and Cat. no. 2138026. Ref. nos. 12138018001 and 12218542001 both include include meters with model number 777.
The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
55,189 meters (for all Recalls Z-1354-1371-06)