AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)
Recall
- Recall Number
- Z-1364-2022
- Event Number
- 90411
- Firm
- Fresenius Medical Care Holdings, Inc.
- FEI Number
- 3001451489
- Product Code
- FIP
- Status
- Terminated
- Root Cause
- Labeling Change Control
- Initiated
- August 3, 2021
- Terminated
- April 10, 2024
- Address
- 920 Winter St, Bld 950, Waltham, MA, 02451-1521
Description
AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)
AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
Customers were notified via telephone beginning 8/3/21. An FMCRTG, LLC Water Systems Specialist contacted the two consignees via phone and scheduled a time to perform the inspection on the AquaBplus 2000 devices. No formal written communication will be sent since calls and inspections have already performed.
US Nationwide distribution in the states of NC and WV.
2 units