FDA Recall Terminated

AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

Recall: Z-1364-2022 · Initiated August 3, 2021

Recall

Recall Number
Z-1364-2022
Event Number
90411
Firm
Fresenius Medical Care Holdings, Inc.
FEI Number
3001451489
Product Code
FIP
Status
Terminated
Root Cause
Labeling Change Control
Initiated
August 3, 2021
Terminated
April 10, 2024
Address
920 Winter St, Bld 950, Waltham, MA, 02451-1521

Description

AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

Reason

AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation

Action

Customers were notified via telephone beginning 8/3/21. An FMCRTG, LLC Water Systems Specialist contacted the two consignees via phone and scheduled a time to perform the inspection on the AquaBplus 2000 devices. No formal written communication will be sent since calls and inspections have already performed.

Distribution

US Nationwide distribution in the states of NC and WV.

Quantity

2 units