FDA Recall Terminated

VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest

Recall: Z-1364-2019 · Initiated March 20, 2019

Recall

Recall Number
Z-1364-2019
Event Number
82561
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
JJE
Status
Terminated
Root Cause
Software design
Initiated
March 20, 2019
Terminated
December 8, 2020
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest

Reason

Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported

Action

Ortho Clinical Diagnostics issued on 20 March 2019, a customer letter via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) customers who were shipped affected VITROS XT 7600 Systems. Software Version 3.4.2 (MOD 4) will be provided for download or DVD format. Contact Ortho Care" Technical Solutions Center at 1-800-421-3311.

Distribution

Worldwide Distribution - US Nationwide in the states of AZ, CA, IL, OH, FL, TX, MI Foreign: Australia, Belgium, Chile, India, Italy, Japan, South Korea, Spain, Canada,, Phillippines

Quantity

85 units: US=19;OUS=66