FDA Recall Terminated

ORTHOFIX, Connector System, LARGE SET SCREW, REF 79-2002

Recall: Z-1364-2018 · Initiated February 23, 2018

Recall

Recall Number
Z-1364-2018
Event Number
79498
Firm
Orthofix, Inc
FEI Number
2183449
Product Code
NKG
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 23, 2018
Terminated
June 18, 2019
Address
3451 Plano Pkwy, Lewisville, TX, 75056-9453

Description

ORTHOFIX, Connector System, LARGE SET SCREW, REF 79-2002

Reason

Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.

Action

The firm disseminated the recall notices on 02/23/2018 by mail. The firm requested the return of the screws.

Distribution

US

Quantity

188 units