FDA Recall Terminated

Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.

Recall: Z-1363-2016 · Initiated January 8, 2016

Recall

Recall Number
Z-1363-2016
Event Number
72894
Firm
Natus Medical Incorporated
FEI Number
3018859
Product Code
FMZ
Status
Terminated
Root Cause
Process control
Initiated
January 8, 2016
Posted
April 11, 2016
Terminated
February 10, 2017
Address
5900 1st Ave S, Seattle, WA, 98108-3248

Description

Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.

Reason

NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.

Action

Natus of Seattle, WA began sending the Urgent Field Safety Notice NatalCare LX Incubator letter (DOC-017939A) to consignees on 1/8/16. A repair kit to correct this issue is now available for repair of affected units. Consignees should complete the NatalCare LX Incubator Oval Port Hinge Repair Kit Order Form and FAX it 847-949-6031 or scan and email this form to [email protected]. Alternately you can telephone Natus Technical Service at 888-496-2887 and order the kit from a specialist. Natus will send the kit(s) free-of-charge to the address you specify. Each upgrade will require up to - 1 hour of time and require simple hand tools. If you have any questions, please contact Technical Service at 888-496-2887 or send email to: [email protected]

Distribution

Distribution US nationwide and Pakistan.

Quantity

75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.