Pressure Monitoring Set, Single-lumen uncoated central venous catheter. RPN C-PMS-250, GPN G02838
Recall
- Recall Number
- Z-1362-2022
- Event Number
- 90393
- Firm
- Cook Incorporated
- FEI Number
- 1820334
- Product Code
- DQY
- Status
- Completed
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 25, 2022
- Address
- 750 N Daniels Way, Bloomington, IN, 47404-9120
Description
Pressure Monitoring Set, Single-lumen uncoated central venous catheter. RPN C-PMS-250, GPN G02838
The products may contain excess coating material on the outside and/or interior of the needle component. Potential adverse events that may occur if an affected product is used include a delay in the procedure, a prolonged procedure, or particulate entering the bloodstream which may result in pulmonary embolism or ischemia.
URGENT: MEDICAL DEVICE RECALL notification letters dated 5/25/22 were sent to customers. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. Action being Taken by Cook Medical Cook Medical is removing devices potentially impacted from the market. An investigation was previously performed for this issue. As a result, actions have been taken to prevent further occurrence of this issue. Lots involved in this recall were manufactured prior to preventative actions. We recognize that this situation is a potential disruption to your normal operations, and we sincerely apologize. Thank you for your immediate assistance in this matter. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. We look forward to your response.
US Nationwide. Brazil, Canada,
11,451 units