FDA Recall Open, Classified

Bio-Rad, REF: 12005660, QXDx BCR-ABL %IS Kit

Recall: Z-1358-2024 · Initiated February 15, 2024

Recall

Recall Number
Z-1358-2024
Event Number
94020
Firm
Bio-Rad Laboratories
FEI Number
3010581187
Product Code
OYX
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
February 15, 2024
Posted
March 22, 2024
Address
5731 W Las Positas Blvd, Pleasanton, CA, 94588-4084

Description

Bio-Rad, REF: 12005660, QXDx BCR-ABL %IS Kit

Reason

Due to incorrect calibrations and controls, their is a potential of incorrect reporting of lower %IS and molecular response values.

Action

On February 7, 2024, Bio-Rad issued their initial "Urgent Medical Device Correction" notification via E-Mail. On February 14, 2023, Bio RAD Laboratories issued a revised "Urgent Medical Device Correction" notification via E-Mail. Bio-Rad asked consignees to take the following actions: 1. Confirm past patient results against correct values for Catalog No. 12005560, Batch Nos. 64562656 and 64571846 when provided with corrected calibrator values by Bio-Rad Laboratories. 2. If all the controls values fall within the corrected ranges, no further action is required. If one or more controls are outside the corrected ranges, that particular run should be considered as failed and the results for the patient samples tested within that run should be considered invalid. 3. Please go to Bio-Rad.com for the updated Certificate of Analysis. 4. Please ensure this notice is passed to all those who need to be aware within your organization or to any organization where the impacted devices have been transferred. 5. Please complete and return the attached response form as soon as possible so that we are assured you have received this important communication. 6. This notice has been reported to the appropriate Regulatory Agencies. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax.

Distribution

US Nationwide distribution in the states of PA, SD, TX, FL, MN.

Quantity

33 units