Stryker Secure 3002, 115 V ~ 60 Hz, 500 lbs ( 227Kg), Stryker, Kalamazoo, MI 49001 Product Usage: Med/Surg beds are AlC powered hospital beds, intended for clinician use, which consist of built-in motors and remote controls that can be operated by the patient and/or clinician to adjust the height and surface contour of the beds. The Secure II model is primarily used in the Med/Surg units, but may also be found in a Critical Care environment.
Recall
- Recall Number
- Z-1354-2012
- Event Number
- 61054
- Firm
- Stryker Medical Division of Stryker Corporation
- FEI Number
- 1831750
- Product Code
- FNL
- Status
- Terminated
- Root Cause
- Component change control
- Initiated
- March 12, 2012
- Posted
- March 30, 2012
- Terminated
- August 29, 2013
- Address
- 3800 East Centre Ave., Portage, MI, 49002
Description
Stryker Secure 3002, 115 V ~ 60 Hz, 500 lbs ( 227Kg), Stryker, Kalamazoo, MI 49001 Product Usage: Med/Surg beds are AlC powered hospital beds, intended for clinician use, which consist of built-in motors and remote controls that can be operated by the patient and/or clinician to adjust the height and surface contour of the beds. The Secure II model is primarily used in the Med/Surg units, but may also be found in a Critical Care environment.
Stryker Medical has determined that units manufactured between September 1, 2004 through November 30, 2007, did not receive updated preventive maintenance information. (Failure to maintain the bed brakes including the brake plate kits, may result in reduced brake holding force. Reduced braking force may result in a patient fall or caregiver muscle strain, either of which may require medical int
Stryker Medical sent an Urgent-Medical Device Correction letter dated March 12, 2012 to all consignees. The letter identified the affected products, problem, updated preventive maintenance information and requested the following immediate actions: 1) Locate the beds listed in this notice, 2) Ensure that the brakes are holding properly, 3) If the brakes are not holding, contact Stryker (1-800-327-0770, option 4) to order the maintenance parts, 3) Continue to periodically inspect brake functionality, as noted on enclosed preventive maintenance schedule/checklist, 4) Order replacement brake components, as needed, 5) Return the enclosed post card to confirm receipt of this notification, 6) If you have loaned or sold any of the beds listed in this letter, please, forward a copy of this notice to the new users and advise us of their new location, 6) If you have disposed of any of the beds and they are no longer in use, please, advise us of their obsolescence by providing us with their serial numbers. For questions or concerns, please, contact Sandy Cliche at 269-389-6025.
Worldwide Distribution - USA (nationwide) including the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and the countries of: DEDEDO GU, HAMILTON ON, SANTIAGO, SHATIN, NT , CAPITOL FEDERAL, ITAPEVI SP, ATHENS, OSAKA, SEOUL, MEXICO DF MX, MONTREUX, BALERNA, NORTH POINT, WARSZAWA, CANOVANAS, KOWLOON BAY, SINGAPORE, NEWBURY, and JOHANNESBURG.