ACCU-CHEK Advantage Blood Glucose Meter/ Blood Glucose Monitoring System; Cat. no. 870. Ref. no. 080087000 with meter model number 870; Ref. no.0353756001 with meter model number 33304510; Ref. no. 03537536001 with meter model no. 2138930; Ref. no. 03537536001 with meter model no. 3288650.
Recall
- Recall Number
- Z-1354-06
- Event Number
- 35022
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- CGA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 22, 2006
- Posted
- August 5, 2006
- Terminated
- April 7, 2008
- Address
- 9115 Hague Rd, Indianapolis, IN, 46250-0416
Description
ACCU-CHEK Advantage Blood Glucose Meter/ Blood Glucose Monitoring System; Cat. no. 870. Ref. no. 080087000 with meter model number 870; Ref. no.0353756001 with meter model number 33304510; Ref. no. 03537536001 with meter model no. 2138930; Ref. no. 03537536001 with meter model no. 3288650.
The meter gives a strip error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error and the manual only instructs the user to retest with a new strip using a larger drop of blood.
A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
55,189 meters (for all Recalls Z-1354-1371-06)