FDA Recall Terminated

Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover) for the Vivo 50 and Vivo 65 ventilators - Product Usage: The protective cover is intended for additional protection of the Vivo 50, Vivo 55, Vivo 60 and Vivo 65 ventilators during transportation, and in hospital, institutional, and home care environments. It can be used while the Vivo ventilators is operating, for example mounted on a wheelchair, in a personal vehicle, or carried by hand. The device is not to be used with the protective cover while positioned in a warm place, such as in direct sunlight or close to a radiator. The protective cover includes the following features: transparent window for accessing front panel and buttons; port for patient circuit, cables, O2 inlet, and standby button; cooling air inlet; patient air inlet; straps for safe mounting; and cooling air outlet.

Recall: Z-1352-2020 · Initiated January 28, 2020

Recall

Recall Number
Z-1352-2020
Event Number
84828
Firm
Breas Medical, Inc.
FEI Number
3010817335
Product Code
NOU
Status
Terminated
Root Cause
Device Design
Initiated
January 28, 2020
Terminated
September 15, 2020
Address
16 Esquire Rd, North Billerica, MA, 01862-2527

Description

Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover) for the Vivo 50 and Vivo 65 ventilators - Product Usage: The protective cover is intended for additional protection of the Vivo 50, Vivo 55, Vivo 60 and Vivo 65 ventilators during transportation, and in hospital, institutional, and home care environments. It can be used while the Vivo ventilators is operating, for example mounted on a wheelchair, in a personal vehicle, or carried by hand. The device is not to be used with the protective cover while positioned in a warm place, such as in direct sunlight or close to a radiator. The protective cover includes the following features: transparent window for accessing front panel and buttons; port for patient circuit, cables, O2 inlet, and standby button; cooling air inlet; patient air inlet; straps for safe mounting; and cooling air outlet.

Reason

The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover.

Action

On January 28, 2020, the firm began distributing Urgent Medical Device Correction letters to affected customers. Customers were informed of the product issue and associated risk of hypercarbia caused by excessive CO2 build-up (rebreathing). Customers were instructed to do the following: " Identify users/patients who use Protective Covers (p/n 006344) with a Vivo 50 or Vivo 65 ventilators. The model of ventilator can be determined from the device label on the bottom cover. Vivo 50 units have part no 215000 and Vivo 65 units have part no 226500. " Check if the Protective Cover is used with an exhalation valve circuit or dual-limb circuit. If this is the case, the users should be advised to take note of Rebreathing alarms and Exhalation Valve Control Error alarms. If any of these alarms occur, the small hole at the back cover of the ventilator may be obstructed. This may be resolved by ensuring that no force is applied to the front or back side of the Protective Cover. " Please notify Breas Medical of your preferred solution from the options below. " Please fill out and return the attached Reply Form at your earliest convenience and latest 29 February 2020. " Please forward this notice to anyone to whom you may have further distributed the product (additional consignees). Breas offers two options/solutions, which can be combined. 1. Advise users of the Protective Covers to be observant of the Rebreathing alarms and Exhalation Valve Control Error alarms and inform them that if these alarms occur, the small hole at the back cover of the ventilator may be obstructed. This can be resolved by ensuring that no force is applied to the front or back side of the Protective Cover. With this precaution, the Protective Covers can continue to be used. 2. If a user prefers, Breas offers to replace Protective Covers (p/n 006344) with new Protective Covers free-of-charge. We can provide you up to the number of the Protective Covers purchased according to our reco

Distribution

Worldwide distribution - US Nationwide distribution in the states of AR, CA, FL, GA, IA, IL, KS, MA, MI, MO, NJ, NY, OH, PR, SC, TX, VA, WI and countries distribution worldwide.

Quantity

629