FDA Recall Open, Classified

Molift 2-Point sling bar, an accessory for Molift Mover 180/Air. Item nos. 1830001 1830002 1830002C 1830003

Recall: Z-1352-2019 · Initiated February 4, 2019

Recall

Recall Number
Z-1352-2019
Event Number
81977
Firm
Moller Vital Hadelandsvegen 2
FEI Number
3004137175
Product Code
FSA
Status
Open, Classified
Root Cause
Use error
Initiated
February 4, 2019
Address
P.o. Box 408, Gjovik Norway

Description

Molift 2-Point sling bar, an accessory for Molift Mover 180/Air. Item nos. 1830001 1830002 1830002C 1830003

Reason

The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall.

Action

On February 4, 2019, the firm notified customers of the recall via Urgent Field Safety Notice letter. The letter alerted customers to the recalled product. The firm stated that all new 2-Point Sling Bars are now equipped with metal alloy hooks instead of the previously used composite hooks. As part of this recall, the firm will replace the recalled devices with the new design. Distributors were asked to do the following: 1) Read the instructions, and complete and return the customer response form 2) Notify affected customers of the recall 3) Review the list of sales records from Etac Supply Gjvik and supply additional information from your records. 4) Identify and contact end-customers/users and decide the requirement for ordering new units 5) Order new 2-Point Sling Bars from Etac Supply Gjvik 6) Ship replacement units to end customers accompanied by replacement instructions (see user manual) and the Consignee Form. The End-User/Customer must perform the replacement and fill the replacement records (green fields) in the consignee form and return a signed copy to the distributor. 7) Receive the signed Consignee Form from end-users/customers. Perform weekly reporting to Etac Supply Gjvik in the same spreadsheet that contains the sales records. The signed Consignee Forms must be returned to Etac Supply Gjvik as a complete package upon finalizing all replacements.

Distribution

Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington, and Wisconsin. Foreign distribution to Canada, Asia, Europe, and Australia.

Quantity

427 (total)