FDA Recall Terminated

Oridion Surestream CO2 Sampling Lines- Sure VentLine H Set Adult/Pediatric Part Number: 010987-03 Product Usage: The intended use of SureStream sampling lines is to convey a sample of a patients expired breath to a capnograph in order to measure CO2 partial pressure in intubated and non intubated patients.

Recall: Z-1350-2012 · Initiated December 1, 2011

Recall

Recall Number
Z-1350-2012
Event Number
61143
Firm
Oridion Medical
FEI Number
3001227743
Product Code
CCK
Status
Terminated
Root Cause
Device Design
Initiated
December 1, 2011
Posted
March 30, 2012
Terminated
June 4, 2012
Address
1987 Ltd. 7 Hamarpe St. P.O. Box 45025, Har Hotzvim Industrial Park Jerusalem Israel

Description

Oridion Surestream CO2 Sampling Lines- Sure VentLine H Set Adult/Pediatric Part Number: 010987-03 Product Usage: The intended use of SureStream sampling lines is to convey a sample of a patients expired breath to a capnograph in order to measure CO2 partial pressure in intubated and non intubated patients.

Reason

Surestream CO2 Sampling Lines-Connector may cause breakage of the microswitch inside the CO2 inlet

Action

Oridion issued a Technical Note communication in March 2011 to field staff and end users globally. The Technical Note identified the affected product and provided instructions for user on the proper insertion of Surestream products. For questions contact your local distributor or Oridion at [email protected]

Distribution

USA Nationwide Distribution including the states of: AZ, CA, FL, HI, LA, MA, MO, NJ, NY, OH, OR, TX and WI.

Quantity

84 boxes (2100 untis)