FDA Recall Open, Classified

Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.

Recall: Z-1349-2023 · Initiated November 11, 2022

Recall

Recall Number
Z-1349-2023
Event Number
91777
Firm
Abbott Molecular, Inc.
FEI Number
3005248192
Product Code
OOI
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
November 11, 2022
Posted
April 6, 2023
Address
1300 E Touhy Ave, Des Plaines, IL, 60018-3315

Description

Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.

Reason

The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This would result in a potential for delay of results and incorrect results.

Action

The field service engineer installed the missing plunger assembly in the clamp bar while at the consignee location on 11/11/2022. The recalling firm issued the recall letter on 3/1/2023 via overnight mail to their sole consignee. The letter explains the issue, the potential impact, and the necessary actions, which includes placing the impacted ADU out of service and contacting Abbott technical support regarding concerns about the performance of the instrument or results. The letter is to be reviewed with the Medical Director or Laboratory Management and the consignee is to follow their laboratory protocol regarding the need for review of previously reported patient results using the affected system. A customer reply form is enclosed with the letter to acknowledge receipt and to indicate it was disseminated to all users, understood, and implemented. If you have any questions regarding this communication, please contact your local Abbott representative or call DVP Quality Assurance, at 224-361-7000.

Distribution

US Distribution to state of: PA, a government account.

Quantity

1