FDA Recall Terminated

COHERENCE Oncologist, Model Nos. 07333680 & 07351898

Recall: Z-1346-2020 · Initiated January 16, 2020

Recall

Recall Number
Z-1346-2020
Event Number
84795
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IYE
Status
Terminated
Root Cause
Software Design Change
Initiated
January 16, 2020
Terminated
January 4, 2021
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

COHERENCE Oncologist, Model Nos. 07333680 & 07351898

Reason

A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.

Action

The firm started notifying customers on January 16, 2020, with letters titled Urgent Medical Device Correction. Customers were advised of the product safety issue. To avoid the possibility of an incorrect alignment of the patient, the user must ensure that filtered images arrive on the COHERENCE Oncologist, or apply any additional filters to the image BEFORE using the "interactive shift" or "landmark registration" tools. The firm will be offering a software solution for the issue free of charge. The firm's letter stated that based on their investigation, customers may continue to use their devices.

Distribution

Domestic distribution to CA, FL, GA, IA, MA, MI, NE, NJ, NV, NY, OH, OK, PA, SD, TN, WV. Worldwide foreign distribution.

Quantity

64 Worldwide/26 Domestic