NAMIC CONVENIENCE KIT MERCY GENERAL HOSPITAL NEURO KIT, REF Catalog No. 601319810, UPN Product No. H7496013198101 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire.
Recall
- Recall Number
- Z-1346-2016
- Event Number
- 72957
- Firm
- AngioDynamics Inc.
- FEI Number
- 3017892510
- Product Code
- OEZ
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 3, 2015
- Posted
- April 7, 2016
- Terminated
- March 16, 2017
- Address
- 10 Glens Falls, Tech Park Glens Falls, NY, 12801-3864
Description
NAMIC CONVENIENCE KIT MERCY GENERAL HOSPITAL NEURO KIT, REF Catalog No. 601319810, UPN Product No. H7496013198101 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire.
Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".
On November 03, 2015, Navilyst Medical distributed Urgent Voluntary Medical Device Recall notices to their customers via courier service. Customers were advised to immediately remove the recall product from inventory and to segregate the products in a secure location for return to AngioDynamics, Inc. Customers should immediately forward a copy of the recall notification letter to all sites if these products were further distributed. If affected product is located in your institution, please call AngioDynamics, Inc. Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday - Friday EST) to obtain a replacement or credit for your returned product. Promptly complete, sign and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return). The Reply Verification Tracking form can be submitted via fax to: Attn: Convenience Kits - Procedure Products Needles 1-800-782-1357 or via email to [email protected] or [email protected].
US Nationwide in the states of CA, IL, MA, SC, TX
45 kits