cobas 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.
Recall
- Recall Number
- Z-1346-2013
- Event Number
- 64930
- Firm
- Roche Molecular Systems, Inc.
- FEI Number
- 2243471
- Product Code
- OOI
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- April 1, 2013
- Posted
- May 21, 2013
- Terminated
- August 7, 2014
- Address
- 1080 Us Highway 202 S, Branchburg, NJ, 08876-3733
Description
cobas 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.
In rare instances, channel shifted results have been generated with cobas 4800 assays run on v1.1.1 of the cobas 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported
Roche sent an "Urgent Medical Device Correction" (UMDC) notification with response form dated April 1, 2013, to all affected customers via UPS Ground on April 1, 2013. The letter identified the product, the problem, and the action to be taken by the customer. A contact number (Roche Diagnostic Technical Support) is provided for technical support 24 hours a day, seven days a week at phone number: 1-888-912-7090.
Nationwide Distribution.
425 instruments