FDA Recall Terminated

cobas 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.

Recall: Z-1346-2013 · Initiated April 1, 2013

Recall

Recall Number
Z-1346-2013
Event Number
64930
Firm
Roche Molecular Systems, Inc.
FEI Number
2243471
Product Code
OOI
Status
Terminated
Root Cause
Software design
Initiated
April 1, 2013
Posted
May 21, 2013
Terminated
August 7, 2014
Address
1080 Us Highway 202 S, Branchburg, NJ, 08876-3733

Description

cobas 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.

Reason

In rare instances, channel shifted results have been generated with cobas 4800 assays run on v1.1.1 of the cobas 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported

Action

Roche sent an "Urgent Medical Device Correction" (UMDC) notification with response form dated April 1, 2013, to all affected customers via UPS Ground on April 1, 2013. The letter identified the product, the problem, and the action to be taken by the customer. A contact number (Roche Diagnostic Technical Support) is provided for technical support 24 hours a day, seven days a week at phone number: 1-888-912-7090.

Distribution

Nationwide Distribution.

Quantity

425 instruments