FDA Recall Open, Classified

CELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood Part Number: 7900001

Recall: Z-1345-2022 · Initiated May 20, 2022

Recall

Recall Number
Z-1345-2022
Event Number
90417
Firm
Menarini Silicon Biosystems
FEI Number
3004153557
Product Code
NQI
Status
Open, Classified
Root Cause
Process control
Initiated
May 20, 2022
Address
3401 Masons Mill Rd, Ste 100, Huntingdon Valley, PA, 19006-3574

Description

CELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood Part Number: 7900001

Reason

High number of total images/unassigned events including (dual positives) and the potential for false positive results being placed into the image gallery in some patient samples

Action

Menarini Silicon Biosystems issued notification telephone Urgent Medical Device Correction Notification CTC Kit Lot S108 to US on 5/20/22. Followed by e-mail direct to consignees for US sites - May 23, 2022 and Coordination with Asia Pacific region and the EU with Local QA and RA from the regions. Letter states reason for recall, health risk and action to take: Discontinue usage of Lot S108 and quarantine the lot in your inventory for disposal. Review patient samples run with Kit Lot S108 which had results that were above the cut-off. If you observe any images with the staining pattern of the false positive images shown below or there is any uncertainty regarding these images impacting the patient's outcome from below the cutoff to above the cut-off, the patient sample should be repeated. Report any previous occurrence of unusually high number of total images/ unassigned events or potential false positive events that have not already been reported to the Customer Technical Support. " Complete and return the enclosed Confirmation of Receipt form no later than May 31, 2022. If you have additional questions about this issue, contact Customer Technical Support at +1-877-837-4339 option 2

Distribution

CA FL IL IN MA MI NC NM NY OH ON PA TX Foreign: Italy, Shanghai, Singapore,

Quantity

484 units