ColorGard, latex-free, Midlength Secondary I.V. Set, convertible pin, 40 inch with Purple striped tubing for use with OMNI-FLOW and PLUM series infusions systems, 15 drops/mL, a sterile Rx single use device, Hospira Inc, Lake Forest, IL USA. Made in Costa Rica, list 4052448.
Recall
- Recall Number
- Z-1344-2010
- Event Number
- 54079
- Firm
- Hospira Inc.
- FEI Number
- 3013319212
- Product Code
- FPA
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- November 24, 2009
- Posted
- April 15, 2010
- Terminated
- January 29, 2012
- Address
- 375 N Field Dr, Lake Forest, IL, 60045-2513
Description
ColorGard, latex-free, Midlength Secondary I.V. Set, convertible pin, 40 inch with Purple striped tubing for use with OMNI-FLOW and PLUM series infusions systems, 15 drops/mL, a sterile Rx single use device, Hospira Inc, Lake Forest, IL USA. Made in Costa Rica, list 4052448.
Reports of difficulty in adjusting the flow rate, which could result in flow rate variances.
Hospira sent the Urgent Device Recall letters dated 11/25/09 to their direct accounts of the affected products and requested these consignees to notify their customers of the recall. The accounts were informed that Hospira ceased manufacturing the I.V. gravity and pump administration sets with the CAIR clamp manufactured by Borla S.P.A. of Italy in March 2008, due to customer reports related to a difference in the zone of control used to adjust the flow compared to the original CAIR clamp design. To avoid the possibility of flow rate variances, Hospira is recalling all remaining sets with the Borla designed CAIR clamp. The accounts were requested to quarantine all inventory of the affected lots, complete the enclosed reply form indicating the amount of product being returned, fax a copy of the completed form to Stericycle at 1-877-377-5114, and include a copy with the returned goods to Stericycle at 2670 Executive Drive, Suite A, Indianapolis, IN 46241.
Worldwide Distribution -- U.S., Puerto Rico, Dominican Republic, Bahamas, Virgin Islands, Argentina, Australia, Bahrain, Belgium, Canada, China, Denmark, Ecuador, EI Salvador, Finland, France, Germany, Greece, Guatemala, Hong Kong, Ireland, Italy, Jordan, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uruguay, and United Arab Emirates.
65,577 sets