FDA Recall Terminated

Boston Scientific Fetch 2 Aspiration Catheter, Catalog No. 109400-001. Product The Fetch 2 Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the peripheral and coronary vasculature.

Recall: Z-1342-2016 · Initiated March 22, 2016

Recall

Recall Number
Z-1342-2016
Event Number
73633
Firm
Boston Scientific Corporation
FEI Number
3002095335
Product Code
DXE
Status
Terminated
Root Cause
Process control
Initiated
March 22, 2016
Posted
April 18, 2016
Terminated
August 11, 2017
Address
2 Scimed Pl, Maple Grove, MN, 55311-1565

Description

Boston Scientific Fetch 2 Aspiration Catheter, Catalog No. 109400-001. Product The Fetch 2 Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the peripheral and coronary vasculature.

Reason

Boston Scientific is recalling all models of Fetch 2 Aspiration Catheter as a result of receiving reports from the field of catheter shaft breaks; including cases requiring the use of a snare or other intervention to retrieve the http://webapps.cdrh.fda.gov/division-tracking/images/trash.pngbroken shaft from the patient.

Action

The firm,Boston Scientific, "Urgent Medical Device Recall Notice - Immediate Action Required" letter dated 3/22/2016. The letter was addressed to Material Manger / Field Action Contact. The letter described the problem and the product involved in the recall and actions to be taken. The customers were instructed to immediately cease further distribution or use of the product and complete and return the Reply Verification Tracking Form via email: [email protected] or fax to: BSC Field Action Center at 1-866-213-1806. Distributors should notify their customers of the recall and inform them that this recall is to the hospital level. For questions regarding this communication, please contact your local Boston Scientific Sales Representative or the Boston Scientific Field Action Center at 763-494-1133 or email: [email protected].

Distribution

Worldwide Distribution - US (nationwide) including DC & VI except DE, ID, SD, and WY and the countries of Australia, Austria, Belgium, Bulgaria, Canada, Chile, Croatia, Cyprus, Ecuador, France, Germany, Great Britain, Hungary, Italy, Japan, Jordan, Korea, Malaysia, Mauritius, Mexico, Netherlands, Philippines, Portugal, South Africa, Spain, Sweden, Thailand, Trinidad, Tobago, and Uruguay.

Quantity

21,155 units