ARCHITECT c8000 Processing Module; List Number 01G06-01. Manufactured by Toshiba Corporation, Medical Systems Division, Japan. Distributed by Abbott Laboratories Diagnostic Division, Irving, Texas
Recall
- Recall Number
- Z-1341-04
- Event Number
- 29587
- Firm
- Abbott Laboratories, Inc
- FEI Number
- 1628664
- Product Code
- CGZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 10, 2004
- Posted
- August 13, 2004
- Terminated
- November 19, 2004
- Address
- 1921 Hurd Drive PO Box 152020, Irving, TX, 75038
Description
ARCHITECT c8000 Processing Module; List Number 01G06-01. Manufactured by Toshiba Corporation, Medical Systems Division, Japan. Distributed by Abbott Laboratories Diagnostic Division, Irving, Texas
Process change in the manufacturing of the AC/DC driver board used on the analyzer may result in one or more of the solid state relays to fail during use.
The firm initiated the recall by Technical Service Bulletin sent to affected customers. The AC/DC boards will be replaced on each analyzer by a firm representative.
Product was distrbued nationwide and to the following foreign countries: Mexico, Australia, Germany, Hong Kong, Canada, Thailand, Grand Cayman, Chile, New Zealand, Columbia, Brazil, and Singapore.
171 units and 25 spare parts distributed