FDA Recall Terminated

Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.

Recall: Z-1340-2013 · Initiated March 18, 2013

Recall

Recall Number
Z-1340-2013
Event Number
64647
Firm
Linvatec Corp. dba ConMed Linvatec
FEI Number
1017294
Product Code
KQM
Status
Terminated
Root Cause
Software design (manufacturing process)
Initiated
March 18, 2013
Posted
May 20, 2013
Terminated
November 6, 2013
Address
11311 Concept Blvd, Largo, FL, 33773-4908

Description

Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.

Reason

The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.

Action

The firm, ConMed, sent an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter dated 3/8/2013 to its customers after initial phone notifications. The letter identified the product, problem, and actions to be taken by the customers.

Distribution

Worldwide Distribution-USA (nationwide) including the Virgin Islands and the states of CA, CT, FL, GA, IN, KY, KS, LA, MD, MI, MN, MO, MS, MT, NE, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, and WI, and the countries of Austria, Belgium, Canada, Denmark, Hong Kong and Zambia.

Quantity

267 units