FDA Recall Terminated

MassCheck Amino Acid, Acylcarnitines Dried Blood Spot Control Level II

Recall: Z-1337-2022 · Initiated May 19, 2022

Recall

Recall Number
Z-1337-2022
Event Number
90266
Firm
Chromsystems Instruments & Chemicals GmbH Am Haag 12 Grafelfing Germany
FEI Number
3003951180
Product Code
JJY
Status
Terminated
Root Cause
Labeling Change Control
Initiated
May 19, 2022
Terminated
May 15, 2024

Description

MassCheck Amino Acid, Acylcarnitines Dried Blood Spot Control Level II

Reason

Too low concentration of glycine was detected in the dried blood controls of the batch 2821 of the products.

Action

The firm issued a recall notification on 05/19/2022 via email. The notification mentions if the consignee has used the affected product, check whether they have determined glycine, replace the old package insert with the updated package insert with release date 18.05.2022, forward a copy of the letter to another laboratory if the affected product was given to them, and submit the response form by 01.06.2022. If users determined glycine, they are asked to clarify with the attending physician or laboratory director, if necessary, whether a back-check of the values for glycine is required.

Distribution

US: NY, OH, CT.

Quantity

1457 units